The Airyx WGA 020 provides long-term stable, real-time measurements of relevant greenhouse gas emissions from wastewater treatment plants along with minimal maintenance and calibration effort.
- Swimming sampling hood with metal cage system. Targeted and stable gas collection directly from the water surface, even in dynamic or harsh environments.
- Stabilised temperature. This ensures high accuracy and minimal drift under fluctuating environmental conditions.
- Continuous concentration measurements. Real-time monitoring of target gases with high temporal resolution for accurate process and emission control.
- Humidity-resistant analysers. Sample gas drying impedes humidity related sensor degradation.
- Heated sampling tube (optional) preventing condensation for loss-free gas transfer between sampling hood and analyser.
Data sheets and specifications are subject to change, and Enviro Technology cannot be held responsible for inaccuracies in manufacturer-issued data sheets. To confirm specifications and obtain the latest issue of the data sheet, please contact us.
Datasheet
Wastewater Gas Analyser WGA 020
Specifications
- Measured Gases: N₂O, CO₂, CH₄, O₂
- Measurement Principle:
- Non-dispersive infrared (NDIR) for N₂O, CO₂, CH₄
- Electrochemical sensor for O₂
- Additional Parameters: Ambient temperature, Ambient pressure, Total sample flow
- Time Resolution: 1 s internal, 1 min average (adjustable)
- Enclosure: Rugged, weatherproof design (IP64 rating)
- Power Supply: 24 VDC, external power supply
- Power Consumption: 30 W nominal, 120 W peak
- Operating Temperature: -10 °C to +40 °C
- Dimensions: 660 × 400 × 240 mm
- Weight:
- Analyser (incl. staffing & sun shield): ~27 kg
- Sampling hood unit: ~30 kg
- Mounting Options: Customer specific
- Data Interfaces: Ethernet, optional 4-20 mA, additional data storage on USB memory stick
Measurement Ranges & Detection Limits
N₂O:
- Range: 0–1000 ppm
- Limit of Detection: 0.5 ppm
- Emission Rate Sensitivity: 0.8 mg/m²/h
CO₂:
- Range: 0–100,000 ppm
- Limit of Detection: 50 ppm
- Emission Rate Sensitivity: 84 mg/m²/h
CH₄:
- Range: 0–1000 ppm
- Limit of Detection: 5 ppm
- Emission Rate Sensitivity: 3 mg/m²/h
O₂:
- Range: 0–100 vol %
- Limit of Detection: 0.1 vol % or 0.5 % of reading
- Emission Rate Sensitivity: n/a