The Model N703U trace-level ozone calibrator is designed for calibration and verification of precision gas analysers.
Combining an integrated ozone generator with a high-accuracy UV photometer, the N703U provides stable and repeatable ozone concentrations across a wide operating range, including ultra-low levels required for modern ambient monitoring applications.
The ozone calibrator uses a photolytic ozone generator and photometer feedback to continuously monitor and control ozone output, ensuring accurate concentration delivery and long-term stability. Adjustable flow rates support both low-level and high-output applications, with flexible automation capabilities for routine calibration tasks.
Instrument functions are managed through a distributed CAN Bus based electronics architecture with independently controlled modules. This design isolates faults to individual modules, simplifying troubleshooting and field service while improving overall system reliability and uptime.
Operation and calibration are managed through the NumaView ™ software interface, with remote access providing full control and access to internal data storage from a PC or tablet.
Applications
- Ambient air monitoring networks.
- Regulatory compliance monitoring programmes.
- Environmental and air-quality research.
- Calibration laboratories.
- Industrial air monitoring applications.
Data sheets and specifications are subject to change, and Cura Terrae Air cannot be held responsible for inaccuracies in manufacturer-issued data sheets. To confirm specifications and obtain the latest issue of the data sheet, please contact us.
Datasheet
Teledyne API N703U Trace-Level Ozone Calibrator
System
- Linearity: ± 1% of full scale
- Precision: 1.0 ppb
- Response time: < 240 seconds to 95%
- Stability (7 days): < 1% with photometer feedback; < 3% without photometer feedback (CNST)
Ozone Generator Module
- Flow rate: 1 to 5 LPM adjustable (with internal zero air source); 1 to 15 LPM adjustable (with external zero air source)
- Maximum concentration: 5 ppm at 1 LPM
- Minimum concentration: 2 ppb at 5 LPM
- Response time: < 240 seconds at 98%
UV Photometer
- Range: 0-100 ppb to 0-10 ppm (selectable)
- Precision: 1.0 ppb
- Linearity: ± 1% of full scale
- Rise/fall time: 20 seconds to 95% (photometer response)
- Response time: < 180 seconds to 95% (system response)
- Zero drift: < 1.0 ppb/7 days
- Span drift: < 1%/24 hours
- Lag time: < 10 seconds
- Flow rate: 800 cc/min ± 10%
Electrical Specifications
- Power requirements: 110V-240V, 50/60 Hz, typical consumption 85W
Communication Specifications
- Included I/O: 1 x Ethernet (TCP/IP), 1 x RS232, 2 x front panel USB device ports
- Optional digital I/O expansion: 10 x control inputs (user definable), 8 x digital status outputs (user definable)
- Optional valve driver expansion: 12 x control outputs capable of driving up to 8 x 8-watt valves (user definable), 1 x analog channel (10V or 5V ranges, user definable)
Physical Specifications
- Operating temperature: 5° to 40°C (41° to 104°F)
- Dimensions: 7″H x 17″W x 24″D (178 x 432 x 609 mm)
- Weight: 31.25 lbs (14.18 kg)