This sampler uses two completely independent sampling lines and provides a constant and precise flow. It provides PM2.5 and PM10 measurements according to the EN12341 standard.
The DIGITEL Low Volume Sampler (LVS) DPA-14 Double Line (DL) is a fully automatic system to sample dust and aerosol particles for later assessment and analysis (gravimetric and analytical determination) in accordance with EN12341.
It is compatible with all standard Digitel DPA-14 inlets.
It features two completely independent sampling lines with automatic filter changers for parallel sampling of PM10 and PM2.5 inlets. The sampler’s operation range in standard execution is 15 to 50 litres per minute (0.9 to 3m³/h). The DIGITEL LVS DPA-14 DL has two magazines of 30 filters each stretched in filter holders for up to 4 weeks of daily sampling of PM10 and PM2.5 samples. They are automatically changed to the flow position at the preset time.
The cabinet housing of the DIGITEL LVS DPA-14 DL is suited for 19’’ rack installation or can be installed on the floor or on a table in a measurement station. An additional noise baffler allows extremely silent operation. Superior workmanship in sampler mechanics backed by the latest technical and electronic control guarantee a long lifetime and absolutely reliable operation.
Data sheets and specifications are subject to change, and Enviro Technology cannot be held responsible for inaccuracies in manufacturer-issued data sheets. To confirm specifications and obtain the latest issue of the data sheet, please contact us.
Datasheet
Digitel DPA-14 DL Aerosol Sampler
Technical specification
Sampling
- Two independent sampling lines
- Flow rate: 15 – 50 l/min
- Filter: round filters of d = 47 mm, flowing area d = 40 mm
- Filter changer capacity: 2 x 30 filters in filter magazines plus 2 x 1 in sampling position
- Application range: -20° to 50° C; 0 % to 95 % RH, -40° to 50° C with additional heater, 2000 amsl
Time programs
- Work, Pause (0 to 99’999 minutes each)
- Start time adjustable, using date and time
- Different sampling cycles programmable
Protocol files
- Data of filter, temperature, pressure, humidity, flow, blower load
- Calibration history, sensor mean value record file
- Settable averaging period 1 min – 24 h
Accuracy
- Constancy of sample flow: <2% with calibration at 20°C, operating at -20°C-+50°C
Operating Data
- Power supply: 230V AC / 50-60 Hz; max. 4A/360 W
- Mean consumption: 160 W 2 coolers (50%): 200 W
- Cooling capacity of compressors: 360 W each (Option), max. power consumption 40W each
- Mean life cycle suction unit: > 16’000 h
- Sensors: Ambient and flow pressure, temperature, humidity, filter storage temperature
Interfaces
- RS232C, USB, Ethernet, RS485
- Interface protocols: DIGITEL, Bayern-Hessen, AK, TCP/IP,HTTP, FTP
- Internal memory: 16MB, ring buffer, filter data of two years of daily sampling
Inlets
- TSP, PM10, PM2.5 and PM1, customized inlets on request
Materials
- Coated aluminium, stainless steel, POM, PTFE, NBR
- Material of sampling line: EMATAL coated aluminium
Dimensions & Weight
- Field housing (without inlet) 526 x 487 x 981 mm, protection class IP54
- 19″ indoor housing (without inlet) 448 x 455 x 756 mm, 36 kg
Noise level
- <50dB(A) at 1m, <32dB(A) at 8m
Sensor specifications
- Flow sensor accuracy (calibrated): < 1 %
- Ambient & internal pressure sensor range/ accuracy (calibrated): 300 – 1100 hPa/ ± 0.12 hPa
- Ambient & internal temperature sensor range/ accuracy (calibrated): -40 – +65°C/ ± 0.5°C
- Ambient & internal humidity sensor range/ accuracy: 0 – 60°C; 0 – 100 %/ ± 3 % rH
- Filter storage temperature sensor range/ accuracy: -20 – +60°C/ ± 0.1°C