The FIDAS 200 is an EN 16450-approved fine dust measurement device for the simultaneous measurement of PM2.5 and PM10 meeting UK reference method ‘equivalency’ certification requirements.
Excellent for regulatory pollution control in monitoring networks. Ideal for ambient air monitoring campaigns, long-term studies, emission source attribution and emission dispersion studies.
- Used extensively in Local Authority and national monitoring networks including the AURN.
- MCERTS Certified: Continuous Ambient Air Monitoring System & MCERTS for UK Particulate Matter
- Type-approved and certified according to latest EN requirements (EN 15267)
- Continuous and simultaneous real-time measurement of multiple PM values
- Additional information on particle number concentration and particle size distribution
- Market leader for continuous and simultaneous monitoring of ambient PM2.5 and PM10 in European and neighbouring countries
- Designed for outdoor installation (with stainless steel weatherproof housing), whereby this does not require full air conditioning, but can only be operated with an auxiliary heater for indoor temperatures below 5°C.
How it works
The Fidas ® 200 fine dust monitor uses the recognised measuring technique of optical light scattering according to ISO 21501-1 on the single particle and is equipped with an LED light source of high light intensity, high light stability and long service life. The instrument’s calibration can be checked easily and quickly, and, if necessary, adjusted at any time, even when installed, using a monodisperse test aerosol.
The sampling system of the Fidas ® 200 operates with a volume flow of approx. 0.3m³/h.
It is equipped with a Sigma-2 sampling head according to VDI 2119, which enables representative sampling even in strong winds, as well as a drying section, which demonstrably prevents falsification of the measurement due to condensation effects at high humidity.
The Fidas ® 200 fine dust monitor offers a wide range of communications options.
Data sheets and specifications are subject to change, and Enviro Technology cannot be held responsible for inaccuracies in manufacturer-issued data sheets. To confirm specifications and obtain the latest issue of the data sheet, please contact us.
Datasheets
Palas Fidas® 200 Particulate Matter Analyser – Datasheet
Palas Fidas® 200 Particulate Matter Analyser – Flyer
Technical specification
- Measuring principle Optical light scattering at single particles
- Reported data PM1, PM5, PM4, PM10, TSP, CN, particle size distribution, ambient pressure, ambient temperature, rel. ambient humidity
- Measurement range (number CN) 0 – 20,000 particles/cm3
- Measurement range (size) 18 – 18 µm (certified range, other measuring ranges on request)
- Measurement range (mass) 0 – 10,000 µg/m3
- Measurement uncertainty7 % for PM2.5, 7.5 % for PM10 (expanded measurement uncertainty according to EN 16450, TÜV Report)
- Volume flow 8l/min =∧0.3m/h3±3%(24h), compliant with EN16450
- Size channels 64 (32/decade)
- Time resolution 1s–24h
- Interfaces USB, Ethernet (LAN), RS-232, Wi-Fi
- User interface Touchscreen, 800 • 480 pixel, 7” (17.78 cm)
- Protocols UIDEP, UDP, ASCII, MODBUS, Bayern-Hessen
- Data logger storage Capacity for 2 years continuous operation at 60 s storage interval
- Software PDAnalyze
- Data acquisition Digital, 20 MHz processor, 256 raw data channels
- Light source Long term stable LED
- Housing Table housing, optional: with mounting brackets for rack-mounting
- Operating system Windows 10 IoT Enterprise
- Power supply 115 – 230 V, 50/60 Hz
- Installation conditions +5 – +40 °C
- Response time < 2 s
- Sampling head Passive collector Sigma-2
- Dimensions 450 • 320 • 180,5 mm (H • B • T), 19”
- Weight Control unit: 9.3 kg, sample head: 2.25 kg, sample tube: 4.5 kg
- Sampling system Drying of the aerosol by IADS (Intelligent Aerosol Drying System)
- Noise emission < 70 dB(A)
- Fuse T2A
- Resolution 1 µg/m3
- Power consumption Normal operation: 60 W, max. 200 W
- Data Management Prepared for connection to the Palas Cloud MyAtmosphere (”MyAtmosphere ready”); internet access and separate registration required. MyAtmosphere terms and conditions of use apply.