GASERA ONE PULSE is a versatile and easy-to-use multi-gas monitoring solution offering accurate ppb-level detection at an industry-leading price. Using Gasera’s patented cantilever enhanced photoacoustic spectroscopy (CEPAS) technology, it can simultaneously measure up to 9 gases plus H₂O, and is configurable across a wide range of environmental, industrial, and occupational safety applications.
The GASERA ONE PULSE uses cantilever enhanced photoacoustic spectroscopy (CEPAS), combining a broadband IR source with narrow band-pass optical filters to achieve ppb-level trace gas detection. Unlike conventional membrane-based photoacoustic sensors, Gasera’s patented cantilever pressure sensor is over 100× more sensitive, enabling miniaturisation without sacrificing performance.
The instrument can be configured for a wide variety of applications including: SF₆ and refrigerant leak detection, anaesthetic gas monitoring, greenhouse gas emissions monitoring, soil gas analysis, occupational health and safety, photocatalysis research, animal husbandry, and fume hood performance testing.
Target gases span anesthetics (desflurane, isoflurane, sevoflurane), greenhouse gases (CF₄, C₂F₆, CO₂, N₂O), hydrocarbons (CH₄, C₂H₂, C₂H₄), inorganics (CO, NO, NO₂, NH₃, SF₆, SO₂), and VOCs (acetone, benzene, ethanol, formaldehyde, toluene, xylenes).
The optional Gasera Multipoint Sampler expands the instrument to monitor up to 12 sampling points automatically and programmably, maximising value per instrument.
Data sheets and specifications are subject to change, and Cura Terrae Air cannot be held responsible for inaccuracies in manufacturer-issued data sheets. To confirm specifications and obtain the latest issue of the data sheet, please contact us.
Datasheet
General
- 19″ 3U housing for tabletop and rack mount operation
- Dimensions: 48.4 cm W x 13.9 cm H x 44 cm D (19.1 in W x 5.5 in H x 17.3 in D)
- Weight: approx 13 kg
- Built-in computer with a 7″ WSVGA display
- Data storage capacity sufficient for at least 1 year of continuous monitoring of a full set of gases with the shortest sampling interval
- Total internal gas volume: 30 ml
- 2 gas connections in the rear including 2 sample input connections equipped with user changeable filters for dust and small particles
- Input voltage: 90…264 Vac, 47…63 Hz
- Input power: 75 W max.
- Interface: Ethernet, USB and optionally Serial over USB, current message or voltage message. Supports MODBUS and AK-protocol
Measurement Specifications
- Response time: 15 seconds to few minutes depending on user configurable channel integration time (C.I.T.) and gas exchange routine
- Detection limit: gas and light source dependent. Typically from sub-ppb to sub-ppm
- Dynamic range: typically 5 orders of magnitude (i.e. 100,000 times the detection limit)
- Repeatability: less than 1% of measured value in operational conditions at the calibration concentration
- Accuracy: limited by the calibration gas accuracy at the calibration concentration. Typically 2–5%
- Temperature stability: ambient temperature change within the operational temperature range will not cause drift
- Pressure stability: sample gas pressure change within the pressure range will not cause drift
Environment
- Operational temperature range: 0°C – +40°C
- Operational humidity: below 90% RH, non-condensing
- Operational pressure range: ambient level
- Dust/water resistance: IP20 (IEC 529)
- Storage temperature range: -20°C – +60°C
- Sample gas temperature: 0 – +49°C
- Sample gas humidity: non-condensing, maximum relative humidity 99% for temperatures up to 30°C, decreasing linearly to 35% relative humidity at 49°C
- Sample gas pressure: 750 mbar – 1050 mbar
- Gas flow: approx 0.6 litres/minute during gas exchange
- Particulates: < 1 µm
Standards